Start Here

Core Concepts Every User Must Know

These ideas apply to almost every instrument you'll touch. Get them straight now and every technology page afterward will make sense.

Bump test vs. full calibration

These two words get mixed up constantly, and the difference matters.

Bump test (a.k.a. functional test)

A bump test is a quick "are you awake?" check. You briefly expose the sensor to a known target gas — above its alarm setpoint — and confirm two things: the reading moves toward the expected value, and the alarm actually goes off (lights, buzzer, vibration). You are not adjusting anything; you're just verifying the instrument responds. It takes under a minute per gas.

Field Tip — bump before every day's use

The industry position (ISEA/OSHA-referenced guidance) is to bump test before each day the instrument is used, and any time you doubt it. A bump proves the whole chain works: sensor, electronics, pump, alarm. It is the single cheapest habit that catches a dead sensor before it kills someone.

Full calibration

A calibration is a two-point adjustment. You zero the instrument in clean air, then expose it to a certified span gas of known concentration and let the instrument reset its internal scale so the reading matches the certified value. Calibration corrects for the slow drift that every sensor experiences as it ages. It takes longer, uses more gas, and is done on a schedule.

✓ Remember — the one-line version

A bump test checks the instrument. A calibration adjusts the instrument. You bump often (daily); you calibrate periodically. A failed bump means calibrate now (or pull the sensor).

Typical calibration schedule

Every technology page lists a calibration & bump schedule specific to that sensor. The numbers above are typical; your manual and SOPs govern.

⚑ Common Rookie Mistake

Thinking "it's calibrated, so I don't need to bump it." Calibration was true on the day it was done. A sensor can die the next morning — from a knock, a poison, or just age — and a calibrated-last-month instrument gives you no warning. The daily bump is what catches that.

Zeroing in fresh air

To zero an instrument is to tell it "this is clean air — call this baseline zero." Most multi-gas monitors auto-zero at power-up, which is exactly why you must turn the instrument on in genuinely clean air — not in the rig with the diesel running, not next to the patient, not downwind of the spill.

⚠ Warning — zeroing in contaminated air

If you power up in air that already contains the target gas, the instrument "learns" that contaminated level as zero. It will then read low by that amount for the rest of the incident — potentially telling you an atmosphere is clean when it isn't. Zero outside, upwind, in known-good air. (Oxygen is the exception you don't want to "zero" — see below.)

Note the oxygen channel isn't zeroed the same way; it's calibrated to the 20.9% of fresh air. Some instruments call the fresh-air startup a "fresh air setup," which sets both the toxic/LEL zeros and the O₂ span at once.

Span gas

Span gas is a cylinder of certified gas at a known concentration used as the "high point" during calibration and bump testing. The concentration is printed on the certificate and the cylinder. A few practical points:

Correction factors & response factors

Some sensors — notably PIDs and catalytic LEL beads — respond to many different chemicals, but they can only be calibrated to one. That calibration gas is the reference. A correction factor (CF), also called a response factor, is the multiplier that converts the reading (in "calibration-gas equivalents") into the true concentration of the chemical you're actually facing.

Analogy: imagine a scale calibrated in apples. If you weigh oranges on it, the number isn't wrong — it's just "in apples." The correction factor is the conversion from apples to oranges.

Field Tip — CFs only work when you KNOW the chemical

A correction factor requires you to already know which single chemical you're measuring. On an unknown, or a mixture, the CF is meaningless — the instrument reports in raw calibration-gas units and you treat the number as relative (rising/falling), not absolute. Use CFs to sharpen a known; never to identify an unknown.

Relative response is the flip side: for a given real concentration, how big a reading does this sensor produce compared to its cal gas? A PID with a low relative response to a chemical (high CF) will under-report it dramatically — some chemicals read only a fraction of their true level, so a small PID number can hide a large hazard.

Sensor lag & response time (T90)

No sensor responds instantly. T90 is the time it takes a sensor to reach 90% of its final reading after being exposed to a step change in gas. Typical T90 values run from a handful of seconds (O₂, CO) to 30+ seconds or more (some toxic and IR sensors). There's also a clear-down (recovery) time — how long to return to zero after the gas is removed, which can be much longer than the rise time for "sticky" gases.

⚠ Warning — walking faster than your meter

If you advance through a space faster than the sensor's response time, you can be well into a hazardous pocket before the reading catches up. Move deliberately, pause at doorways and low/high points, and give the instrument time to settle before you trust a "clean" reading. The alarm that saves you is only useful if the gas reaches the sensor before you reach the gas.

Sampling: pumps vs. diffusion

How does the gas get to the sensor? Two ways:

⚑ Common Rookie Mistake

Forgetting the travel time down a long sample line. If you're drawing through 25 feet of tubing, the gas at the probe tip takes seconds to reach the sensor. A common rule is to allow roughly 1–2 seconds per foot of tubing (check your pump's spec) and then let the reading stabilize before you move the probe. Also do a pump/flow check — a kinked line or blocked filter means you're sampling nothing.

Sample tubing & "sticky" chemicals

The tubing isn't neutral. Some chemicals adsorb onto (stick to) the walls of the sample line, so the sensor reads low or delayed until the tubing "saturates," and then reads high on clear-down. Water-soluble and polar gases — hydrogen fluoride (HF), ammonia, chlorine, HCl — are notorious for this.

Field Tip

HF is the classic teaching case: it clings to ordinary tubing and to the instrument's inlet path, so a pumped reading can badly under-represent the true concentration. For suspected HF, lean on colorimetric tubes and fluoride paper as cross-checks and treat gas-monitor numbers with suspicion.

Datalogging

Most modern monitors continuously record readings to internal memory (a datalog) with timestamps. This matters after the incident: it documents responder exposure, reconstructs what the atmosphere did over time, supports rehab/medical decisions, and provides a record for the AHJ. Know how your instrument's datalog is downloaded and make sure clocks are set correctly — a datalog with the wrong time is nearly useless for exposure reconstruction.

Alarm setpoints

Setpoints are the concentrations at which the instrument alarms. Most gases have two (a low and a high), and toxics often add exposure-based alarms:

Setpoints are usually shipped at sensible defaults, but they can be changed — which is a double-edged sword. Confirm your fleet's setpoints match your SOPs, and never quietly lower an alarm just because it's "annoying."

Exposure limit definitions

These terms turn a raw ppm number into a decision. They come from different organizations and mean different things — mixing them up leads to bad calls.

Occupational and emergency exposure terms you'll hear on scene.
TermStands forPlain-language meaning
PELPermissible Exposure LimitOSHA's legally enforceable limit for worker exposure, usually an 8-hour average. The regulatory floor.
TLVThreshold Limit ValueACGIH's recommended guideline (not law). Often more current/protective than the PEL.
TLV-TWATLV – Time-Weighted AverageAverage concentration for a normal 8-hour day / 40-hour week that most workers can tolerate without ill effect.
STELShort-Term Exposure LimitMax average over a short window (typically 15 min) you shouldn't exceed, even if the 8-hour average is fine.
Ceiling (C)Ceiling limitA concentration that should never be exceeded at any instant — no averaging allowed.
IDLHImmediately Dangerous to Life or HealthThe "get out now / max-protection only" level: exposure that could cause death, permanent harm, or prevent self-escape. NIOSH-published.
ERPGEmergency Response Planning GuidelinesCommunity/emergency tiers (ERPG-1/2/3) for 1-hour exposures — used for public protective actions and evacuation planning.
AEGLAcute Exposure Guideline LevelsEPA emergency tiers (AEGL-1/2/3) across multiple exposure durations (10 min to 8 hr) — the modern successor set for public emergencies.
✓ Remember

Roughly increasing severity for a responder: TWA (all-day OK) → STEL / Ceiling (short bursts, don't linger) → IDLH (immediate threat, maximum protection or get out). Match the number your meter shows to the right limit before you act — a reading below the IDLH can still be well above a limit that matters for a long operation.

Full one-line definitions of every acronym live in the Glossary. Next, start on the sensors themselves: Electrochemical Sensors →